624074/07 F-35
F.8.1.3 Enabling and configuring monitoring settings
To enable and configure monitoring settings:
1. Open the
System ->
Sensors
on/off
window.
Figure F-8. Selecting sensor options
2. Select the check box for the appropriate SpO
2
option(s) to
enable. If two sensors are used, select a
Master
option.
If you specify a sensor to use as the master, that device’s
data is used when the information between the two differs.
By default, the left sensor is selected. If using two SpO
2
sensors,
Mixed
mode ensures INTELLiVENT
®
-ASV uses the
raw data and quality index to determine which sensor to
use as the master sensor.
3. Select the
SpO2
tab on the left.
Figure F-9. Configuring Masimo monitoring settings