624074/07 5-15
This action sounds a medium-priority alarm, displays Apnea
ventilation, and provides ventilation as described in this table:
If the patient triggers two consecutive breaths, the device
reverts to ventilation in the original support mode and at the
original settings, and it displays
Apnea ventilation ended
.
After apnea backup ventilation is enabled or disabled, it retains
this status in all applicable modes. Apnea backup ventilation
requires no clinician intervention, although you can freely
change the mode during apnea backup ventilation, either
switching to a new mode or accepting the backup mode as
the new mode.
When apnea backup ventilation is disabled, the high-priority
alarm message
Apnea
is displayed when apnea occurs.
To verify a setting or to change backup control settings, open
the Controls window.
5.6 Additions window
This section discusses the Additions window and the choices
you can select. You can enable or disable the sigh function, set
tube resistance compensation (TRC) or use the cuff pressure
controller, IntelliCuff
®
, function from the Additions window.
For detailed information on the cuff pressure function, see
Section 3.5.
If the original
support mode
is...
the HAMILTON-
G5 enters this
backup mode
and ventilates
using these
settings
SIMV (S)CMV Control settings
shown in the
Con-
trols
window. (Set-
tings shown are from
original mode (where
possible), or they are
standard settings.)
SPONT, P-SIMV,
DuoPAP, APRV, NIV
P-CMV
APVsimv APVcmv
NIV-ST P-CMV

Table of Contents