A-42 624074/07
A Specifications
A.12 EMC declarations (IEC 60601-1-2)
The HAMILTON-G5 ventilator is intended for use in the electro-
magnetic environment specified in Table A-14, Table A-15,
and Table A-16. The customer (or the operator) of the HAMIL-
TON-G5 ventilator should ensure that it is used in such an envi-
ronment.
The HAMILTON-G5 is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled.
The customer or the user of the HAMILTON-G5 can help pre-
vent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications
equipment (transmitters) and the HAMILTON-G5 as recom-
mended in Table A-16, according to the maximum output
power of the communications equipment.
Table A-14. Guidance and manufacturer's declaration – electro-
magnetic emissions
Emissions test Compli-
ance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The HAMILTON-G5 ventilator uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class A The HAMILTON-G5 ventilator is suitable
for use in all establishments other than
domestic and those directly connected to
the public low-voltage power supply net-
work that supplies buildings for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Com-
plies