A-26 624074/07
A Specifications
A.7 Alarms
Table A-10 is an alphabetical list of the ventilator alarm
settings, their ranges, plus the automatic and standard
alarm settings.
Table A-9. Real-time curves
Parameter Range
Volume (V) 0 to 3200 ml
Flow -200 to 200 l/min
Airway pressure (Paw) -10 to 120 cmH2O
Auxiliary pressure (Paux) -10 to 120 cmH2O
Time 5 to 60 s
FCO
2
0 to 10%
PCO
2
0 to 100 mmHg
0 to 14 kPa
Table A-10. Adjustable alarms with automatic and standard settings
Parameter Unit Range Standard
setting
Automatic
setting
Resolu-
tion
ADULT
ExpMinVol
high
l/min Off,
2-50
10 - ASV:
(IBW*0.1+
40%)*%Min-
Vol, if >100%
- All other
modes:
ExpMinVol +
40%
1