A-6 624074/07
A Specifications
A.5 Control settings
Table A-5 is an alphabetical list of the HAMILTON-G5’s control
settings, their ranges, standard settings, and resolutions.
Table A-6 lists the control settings that apply to the various
ventilation modes.
Table A-7 lists the P/V Tool setting ranges, standard settings,
and resolutions.
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution
Backup Enabled,
disabled
Enabled --
ETS (expiratory trigger sensitivity)
Adult 5 to 70% (of
inspiratory peak
flow)
25% 5%
Pediatric/infant/
neonatal
5 to 70% (of
inspiratory peak
flow)
15% 5%
Flow Pattern
Adult/pediatric Sine, square,
100% decelerat-
ing, 50% decel-
erating
50% deceler-
ating
--
Infant/neonatal -- -- --