624074/07 A-5
Mains fuses (2) T 5.0 AH 250 V
Batteries Electrical specifications:
12 V dc, 15 Ah (internal battery)
14.4 V dc, 6.6 Ah (extended battery pack)
Type: Sealed lead-acid (internal battery) and lithium ion
(extended battery pack), supplied by HAMILTON MEDICAL
only
Operating time per internal battery or extended battery pack
(at standard settings and with nebulizer and communications
interface option enabled): 1 hour minimum. These operating
times apply to new, fully charged batteries not exposed to
extreme temperatures. The actual operating time depends
on battery age and on how the batteries are used and
recharged.
Recharge time: 15 hours minimum (internal battery) and
7 hours minimum (extended battery pack) while ventilator
is connected to ac power
Storage: Remove extended battery pack from ventilator for
long-term storage. Store at -20 to 40 °C (-4 to 104 °F), but
preferably below 30 °C (86 °F), with relative humidity of
25 to 85%. Storage place must be free from vibration, dust,
direct sunlight, and moisture. Avoid storing battery packs for
more than 12 months. To maintain battery integrity, recharge
the batteries at these intervals:
Storage temperature
< 20 °C (68 °F)
20 to 30 °C (68 to 86 °F)
30 to 40 °C (86 to 104 °F)
Recharge interval
Every 9 months
Every 6 months
Every 3 months
NOTE:
Battery life specifiations are approximate. The actual
battery life depends on ventilator settings, battery
age, and level of battery charge. To ensure maximum
battery life, maintain a full charge and minimize the
number of complete discharges.
Table A-4. Electrical specifications