3.6.6 Alarms
When an alarming condition occurs on the humidifier, the
alarm is displayed in the following locations: both on the
humidifier and in the ventilator alarm area.
On the humidifier display
Ventilator alarm area
System -> Humidifier
window on the ventilator
To silence the alarm, press the Alarm Silence key on the humid-
ifier or on the ventilator.
Table 3-3 lists the humidifier-related alarms.
Table 3-3. Humidifier-related alarms
Alarm Do this ...
High water level in the
humidifier chamber
Possible causes: The autofeed mechanism is defective,
the pressure in the water reservoir is too high or
humidifier is mounted at an excessive angle.
Empty humidifier to reduce water level.
Replace the humidifier chamber.
Check the mounting of the humidifier to avoid
excessive angles.
Low water level in the
humidifier chamber
Possible causes: Empty water reservoir, water level
below minimum, tilted humidifier, defective autofeed
mechanism or no chamber inserted.
Insert humidifier chamber and check water reser-
voir.
Refill empty humidifier chamber.
If the water reservoir is empty, connect a new
water reservoir.
Check the mounting of the humidifier to avoid
excessive angles.