3-46 624074/07
3 Hardware options
Temperature too high
near the patient
Possible cause: Ambient temperature too high.
Check whether the circuits are covered by the
patient’s bed covers.
Check whether the circuit or the humidifier
chamber is directly exposed to sunlight.
Check whether the temperature settings are
correct.
Temperature too low/
target humidity not
reached
Possible causes: The device was just turned on,
ambient conditions outside recommended range or
flow beyond specified range.
Wait until the system heats up completely (up to
30 minutes).
Check whether all settings are correct.
Avoid direct air flow from air conditions, etc. to the
humidifier and breathing circuits.
Humidifier: Check
(left or right) tube +
orange light on
connection indicator
on humidifier (item 3
in Figure 3-29)
Check the breathing circuit connection on the speci-
fied side. A green light indicates a good connection.
Humidifier: Red light
on connection
indicator on
humidifier (item 3 in
Figure 3-29)
Service the humidifier. See the HAMILTON-900 Operat-
ing Instructions and the Service Manual.
Table 3-3. Humidifier-related alarms
Alarm Do this ...