3-12 624074/07
3 Hardware options
3. When you insert the sample cell into the receptacle the
sampling pump automatically turns ON. Removal of the
sample cell turns the sample pump OFF.
4. Install the airway adapter between flow sensor and ET tube.
5. The sensor cable must face away from the patient. Safely
secure the sensor cable out of the way, to do so, attach the
sensor cable holding clips to the airway tubing, and then
connect the sensor cable to the clips.
Figure 3-8. Attaching the CO
2
sensor to the airway adapter
1 CO
2
sensor
2 Adapter
To remove the sampling kit sample cell from the sample cell
receptacle, press down on the locking tab and pull the sample
cell away from the sample cell receptacle.