EasyManua.ls Logo

Hamilton G5 - Others

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-6 624074/07
6 Neonatal ventilation
When flow triggering is active, at the late stage of exhalation
the ventilator delivers a constant base flow from the inspiratory
limb to the expiratory limb. The base flow is minimally 1 l/min
and can be as high as 30 l/min, depending on what is the
Flowtrigger sensitivity setting. If the leak flow is higher than the
Flowtrigger setting, autotriggering occurs. To solve the prob-
lem, you can raise the Flowtrigger setting (that is, decrease the
sensitivity) until the autotriggering stops (for details, see
Appendix D).
Leakages are compensated automatically over IntelliTrig (see
Appendix D).
6.6.3 P-ramp
If a neonatal patient has stiff lungs (for example, RDS), be care-
ful when using a short P-ramp (pressure rise time). A very short
P-ramp in this case may cause pressure overshoot.
6.7 Others
WARNING
Prolonged exposure to high oxygen concentra-
tions could cause irreversible blindness and pul-
monary fibrosis in pre-term infants.
High rate settings, very short TI or TE can cause
incomplete inspiration or expiration.
NOTE:
For integrated Aeroneb
®
nebulization of neonatal
patients, see Section 3.4.
Pneumatic nebulization is disabled in the infant/neonatal
application.
During the O
2
enrichment maneuver the applied oxygen
concentration is increased by 25% of the last oxygen setting
(for example: last oxygen setting 40%, resulting oxygen
concentration during O
2
enrichment maneuver 50%).

Table of Contents

Related product manuals