The actual applied oxygen concentration is displayed on the
oxygen control knob (green). Oxygen enrichment continues for
2 minutes unless you terminate it by pressing the O
2
enrich-
ment key again, or manually activating and confirming the
oxygen control knob
The maximum inspiratory or expiratory breath hold time in
infant/neonatal modes is 3 seconds, compared to 10 seconds
in the adult or pediatric modes.