7-2 624074/07
7 Monitoring
7.1 Introduction
WARNING
The device’s oxygen monitoring function can be
disabled in the system window. To prevent
possible patient injury due to nonfunctional
alarms and monitoring, however, HAMILTON-
MEDICAL recommends that oxygen monitoring
always be enabled.
In case of malfunction of the ventilator’s built-
in monitoring and to maintain an adequate
level of patient monitoring at all times, it is
mandatory that additional independent moni-
toring devices be used. Regardless, it always
remains the full responsibility of the operator
to ensure proper ventilation and patient safety
in all situations.
NOTE:
To ensure that oxygen monitoring is always fully
functional, replace an exhausted or missing oxygen
cell as soon as possible or use an external monitor
that complies with EN ISO 21647.
Dashes displayed in place of monitored data indicate
that valid values are not yet available, or do not
apply.
Selections with a gray background are not active and
cannot be selected.
If a ventilator control does not respond when
selected by touch, or by the turn of a P&T knob, the
control is not active in this particular instance or the
function is not implemented.