EasyManua.ls Logo

Hamilton G5 - Operating Modes; Cuff Pressure Control off; User Interface for Automatic Cuff Pressure Control (Manual or no Automatic Adjustment Activated)

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 3-35
3. Then open the
System -> IntelliCuff
window, make
necessary adjustments, and activate cuff pressure adjust-
ments.
3.5.6 Operating modes
The automatic cuff pressure controller works in different
modes of operation allowing the operator to find the suitable
solutions for most clinical situations.
3.5.6.1 Cuff pressure control OFF
CAUTION
The inflating tube will remain initially pressurized
and will not indicate that the automatic adjust-
ment is deactivated and switched OFF. When the
cuff pressure control is switched OFF then all
alarms related to the automatic cuff pressure
controller are disabled.
Selecting the
OFF
button deactivates automatic cuff pressure
adjustments. When OFF, the cuff pressure is not released, how-
ever, leakage of the cuff will no longer be compensated.
Figure 3-24. User Interface for automatic cuff pressure
control (manual or no automatic adjustment activated)