624074/07 F-17
For information on where the SpHb and SpOC data is
displayed, see Section F. 4. For details on performing spot check
updates, see Section F. 5.
F.2.2 About Nihon-Kohden
The Nihon-Kohden SpO
2
pulse oximeter takes a reading every
second to provide accurate, reliable data for SpO
2
and heart
rate (pulse).
The values of these parameters are integrated into the ventila-
tor’s MMP display, support trend graphics, and are subject to
applicable alarms, all of which are controlled at the ventilator.
F.3 Monitored parameters and settings
Table F-3 lists the pulse oximeter-related parameters that are
recorded.
NOTE:
• When a parameter shows dashes or no value, it is
not used in any calculations.
• When running the SpHb sensor in Spot Check mode,
the SpHb and SpOC values are displayed on the
screen only for 1 minute after obtaining the reading
(SpHb) and making the calculation (SpOC).
• HLI is only available with the Nihon-Kohden pulse
oximeter.
Table F-3. SpO
2
parameters and settings
Setting Description Display
Range
Resolution
SpO
2
(%)
1
Nihon-Kohden
Arterial oxygen saturation
in blood
70 to 100 1
SpO
2
(%)
2
Masimo
Arterial oxygen saturation
in blood
0 to 100 1