F-18 624074/07
F Pulse oximetry
HLI Nihon-Kohden only. Heart-
Lung Interaction Index. For
details, see Appendix D.
-200 to 200 1
Pulse Rate (bpm)
1
Heart rate 0 to 240 1
Perfusion index
(%)
2
Pulse strength 0.00 to
20.0
.01 if value
<10
.1 if value
>= 10
SpCO (%)
3
Carbon monoxide
concentration in arterial
blood.
Available in Continuous or
Spot check mode. See Sec-
tion F.2.1.1.
0 to 100 1
SpHb (g/dL)
3
Total haemoglobin in
arterial blood
Available in Continuous or
Spot check mode. See Sec-
tion F.2.1.1.
0.0 to 25.0 Configurable:
0.1 (default)
0.5
1
SpHb (mmol/L)
3
Total haemoglobin in
arterial blood
Available in Continuous or
Spot check mode. See Sec-
tion F.2.1.1.
0.0 to 16.0 Configurable:
0.1 (default)
0.5
1
SpMet (%)
3
Methaemoglobin concen-
tration in arterial blood
0.0 to
100.0
0.1
SpOC (mL/dL)
3
Calculated measurement
of amount of oxygen in
arterial blood
0 to 40 1
1. Both Nihon-Kohden and Masimo
2. Masimo only
3. Available only with the Masimo Rainbow SET option. Not compatible with Nihon-Kohden.
Table F-3. SpO
2
parameters and settings