C.6 ASV Start up
When ASV is started, the device delivers 3 (three) test breaths
in the synchronized intermittent mandatory pressure ventila-
tion mode. The device automatically selects the values for SIMV
rate, inspiratory time (TI), and inspiratory pressure (Pinsp) based
on the calculated IBW, which is determined from the operator-
set Patient height and Gender settings, and according to infor-
mation described in Table C-6 and Table C-7.
Table C-6. Initial breath pattern for Adult settings
IBW
(kg)
P insp
(cmH2O)
TI (s) SIMV
rate
(b/min)
Minimum
target rate
(b/min)
30 to 39 15 1 14 7
40 to 59 15 1 12 6
60 to 89 15 1 10 5
90 to 99 18 1.5 10 5
> 100 20 1.5 10 5
Table C-7. Initial breath pattern for Pediatric settings
IBW
(kg)
P insp
(cmH2O)
TI (s) SIMV
rate
(b/min)
Minimum
target rate
(b/min)
3 to 5 15 0.4 30 15
6 to 8 15 0.6 25 12
9 to 11 15 0.6 20 10
12 to 14 15 0.7 20 10
15 to 20 15 0.8 20 10
21 to 23 15 0.9 15 7
24 to 29 15 1 15 7