F.5 Working in SpHb Spot Check mode
NOTE:
SpHb monitoring is a Masimo Rainbow SET option
(Section F.2 .1 ), and must be installed, enabled, and con-
figured for Spot Check (Section F.8.1.1).
In Spot Check mode:
SpHb and SpOC values are only displayed in the
Monitoring window, SMP, and MMP for 1 minute, as
this data is variable. After 1 minute, the last
measurement value is visible in the
Tools
->
Spot
Check
window. See Figure F-4.
Trend graphs show a single point.
Any alarms related to SpHb or SpOC are only active
for 1 minute. Alarm history is available in the Event
log.
In this mode, SpHb measurements are only recorded when you
manually initiate a measurement. The SpOC value is calculated
using the SpHb measurement as one of its inputs.
As described in Table F-2, the SpHb sensor offers several opti-
ons, including 66 hours (approx. 400 spot checks) of opera-
tion. For each spot check requested, you have an additional
Free Retry period during which you can request repeat mea-
surements. A progress bar shows you the remaining available
time.
Each set of measurements takes approximately 1 minute to
read. If the sensor cannot get a good reading in this time, it
continues to attempt to measure these values until the Free
Retry period is up. If the sensor cannot capture a valid mea-
surement, the Spot Check attempt is not counted and the time
is not subtracted from the total.