624074/07 2-9
NOTE:
Once installed, you can configure and monitor the
HAMILTON-H900 humidifier data and main settings
directly on the ventilator screen, in the
System
->
Humidifier
window. You can also access the
Humidifier window using the Quick Access button
on the main screen. For details, see Section 3.6.
The integrated humidifier is not available in all
markets. Check with your distributor to determine
availability.
Before proceeding with installation, be sure to read
the HAMILTON-H900 Operating Instructions, and the
installation/configuration information in this section
and in Chapter 3, Hardware Options.
Installation involves the following steps:
1. Connecting the power cord, and the communication cable
between the humidifier and the ventilator. See Section
2.4.1.
2. Installing the humidifier onto a mounting bracket. See Sec-
tion 2.4.1.
3. Configuring the humidifier. See the HAMILTON-H900
Operating Instructions.
For details on the integration between the HAMILTON-H900
humidifier and the ventilator, see Section 3.6.
2.4.1 Connecting and attaching the humidifier
NOTE:
When connecting the HAMILTON-H900 humidifier to
the optional internal power socket on the ventilator, be
sure to connect the grounding cable to the potential
equalization terminal on the back of the ventilator and
to a grounding connection in the ICU.