624074/07 F-29
F.7 Masimo information codes and troubleshoot-
ing
The
System -> Info 2
tab shows detailed sensor informa-
tion. This section describes what is displayed. For trouble-
shooting details, see Section F. 10 .
Be sure all desired alarms are enabled (Section F.8.1.5) and cor-
rectly set. For alarm settings, see Table F-7.
To display the window:
1. Open the
System -> Info
window.
2. Touch the right arrow in the top left corner to display page
2. See Figure F-6.
Information is displayed for the attached sensor(s).
SpOC ml/dL 0 - 40
Perfusion Index
(PI) high
% Low: 0.03 - 18 (default: -- off)
• 0.03 - 0.10, in increments
of 0.01
• 0.10 - 1.0, in increments of
0.10
• 1.0 - 18, in increments of 1
High: 0.04 - 19 (default: -- off)
• 0.04 - 0.10, in increments
of 0.01
• 0.10 - 1.0, in increments of
0.10
• 1.0 - 19, in increments of 1
Table F-7. Adjustable alarms with automatic
and standard settings
Parameter Unit Range
For all settings, -- indicates the alarm is off.