F.8.1.5 Enabling and configuring Masimo alarms
The alarms you can configure depend on which options you
have installed. Unavailable alarms are ghosted in the Configu-
ration tab.
For a description of the alarms, see Section F. 6. To specify an
alarm delay for audible alarms, use the
System -> SpO2
tab
as described in Section F. 8 .1 .3.
To enable individual alarms:
1. Open
Configuration
->
Customize
. See Figure F-10.
2. Select the alarms to enable, then select
Close
and
Close/
Save
.
Sensitivity
mode
Specifies the sensor sensitivity,
which can be tailored to different
patient conditions.
Maximum. For
patients with low
perfusion.
Normal (default)
APOD. (Adaptive
sensor off detec-
tion) For cases
where it is likely
that the sensor may
be dislodged.
FastSat Provides quick SpO
2
sampling and
display. May show more changes in
rate, as it is not an averaged value.
On
Off (default)
Table F-11. Masimo monitoring parameters