624074/07 3-29
2. To stop the nebulizer at any time, press the Nebulizer key
on the cockpit. The green backlight changes on the key to
white indicating the nebulization has stopped.
3. To begin the continuous nebulization cycle: connect the
luer adaptor to the filler port, connect the medication sup-
ply to the luer fitting and press the
Nebulizer
key on the
cockpit.
During active nebulization the key is backlit (green) show-
ing the nebulization cycle is in progress. In
Systems ->
Nebulizer
a timer indicates the elapsed nebulization time.
4. To stop the nebulizer at any time, press the
Nebulizer
key
on the cockpit. The green backlight on the key changes to
white indicating the nebulization has stopped.
3.4.10 Functional test
Perform a functional test of the Aeroneb System before first
use (or at any time to verify proper operation). Complete the
test using these steps:
1. Visually inspect each part of the System for cracks or dam-
age. Remove and replace parts if any defects are visible.
2. Pour 1-6 ml of sterile water or normal saline into the nebu-
lizer unit.
3. Connect the nebulizer unit to the control module using the
control module cable.
4. Select
System -> Nebulization
. Start Aerogen as
active nebulizer.
5. Press the
Nebulizer
key on the Ventilation Cockpit and
verify that the key is illuminated green and that aerosol
activity is visible.
6. Press the
Nebulizer
key again to turn the System OFF and
verify that the indicator backlight turns white to indicate
that nebulization has stopped.