624074/07 12-7
NOTE:
To determine whether nebulization is enabled, view
System -> Info
window. If it is not enabled,
enable it in
System -> Nebulization
.
Pneumatic nebulization is inactive if the nebulizer
volume is greater than 50% of the total delivered
volume.
For infant/neonatal ventilation pneumatic nebuliza-
tion is inactive.
Delivered ventilation is compensated for the contri-
bution of the internal nebulizer so that the expected
volume and pressure are delivered.
For more information, see Section 3.4.
The device’s pneumatic nebulization function provides nebuli-
zation during the breath phases and for the duration defined
during configuration. To start nebulization, press the
Nebu-
lizer ON/OFF
key. This function can be activated in all
modes of ventilation. To terminate nebulization before the set
time, press the key a second time.
For effective nebulization, use a pneumatic nebulizer jar tested
by HAMILTON MEDICAL (see Table 1-1). Sections 2.5 and 3.3
describe how to install the nebulizer. The Aeroneb
®
nebulizer is
integrated (see Section 3.4).
12.6 Print screen
NOTE:
The use of HAMILTON MEDICAL CompactFlash
®
or USB
is recommended.
The print screen function saves a JPG file of the current ventila-
tor screen to a CompactFlash
®
or USB storage device. To create
a screen shot, insert one of these storage devices into the
CompactFlash
®
connector, or insert an USB memory stick into
the USB connector; then press the
Print screen
key while
the desired screen is shown.