A-38 624074/07
A Specifications
A.10 Other technical data
Table A-13 lists other ventilator technical data.
Table A-13. Other technical data
Parameter Specification
Inspiratory time (SPONT
breaths)
0.2 to 3 s
0.2 to 10 s (ASV)
Minimum expiratory time 20% of cycle time
Automatic expiratory base flow Pressuretrigger = 1l/min
Flowtrigger setting
≤
1l/min: 2l/min
Flowtrigger setting > 1l/min: 2*Flowtrigger
Maximum limited pressure 120 cmH
2
O, ensured by overpressure valve
Maximum working pressure 120 cmH
2
O, ensured by high Pressure limit
Measuring and display devices Pressure and volume measurements:
Type: Differential pressure transducer,
variable orifice
Sensing position: Patient Y-piece
Measurements: See Table A-8
Time measurements:
Type: Microprocessor
Sensing position: Inside ventilator
Measurements: See Table A-8