624074/07 2-21
Figure 2-16. Installing the expiratory valve cover
WARNING
To prevent inaccurate flow sensor readings, make
sure the flow sensor is correctly installed:
The flow sensor tubings must not be bent.
The flow sensor tubings must be secured with
clamp (included with flow sensor).
Use HAMILTON MEDICAL flow sensors only.
Flow sensor: Insert a flow sensor for the proper patient type
between the Y-piece of the breathing circuit and the patient
connection (Figure 2-17). Connect the blue and colorless tubes
to the flow sensor connectors in the front panel. The blue tube
goes to the blue connector. The colorless tube goes to the
silver connector.