624074/07 3-33
is installed. The connector will work on only one end of the
disposable tubing provided by HAMILTON MEDICAL.
2. On the ventilator side of the disposable tubing a shut-off
valve is built into a luer connector. The shut-off valve pre-
vents loss of cuff pressure in case of a disconnect of the
tubing from the ventilator. The luer connector fits on the
ventilator only at the automatic cuff pressure controller
port.
3. The patient side of the disposable tubing is designed to fit
the connector (pilot balloon) for cuff pressure measurement
on the ET tube or the tracheotomy cannula.
Inspect all parts before use, and do not use if parts are missing,
cracked or damaged. If there are missing parts, malfunction
or damage, contact your HAMILTON MEDICAL product sales
representative for replacements. Use only with components
specified by HAMILTON MEDICAL.
3.5.4 Enabling the integrated cuff pressure controller
option
1. To operate the cuff controller select
Configuration ->
Options
.
2. Select
IntelliCuff
to activate the controller.