J.5 Quick Wean: Choosing the default weaning
activity criteria
Only available with the INTELLiVENT
®
-ASV option. The Quick
Wean configuration screen gives you access to the settings to
specify conditions for starting weaning activities, settings to be
maintained during an SBT, and conditions for aborting ongoing
weaning activities.
The factory default settings are evidence based (see references
in the INTELLiVENT
®
-ASV manual). However, you can change
the settings if you use a different protocol.
For detailed information about setting the parameters, recom-
mended settings, and using Quick Wean, see the INTEL-
LiVENT
®
-ASV manual.
Table J-1. Timing controls applicable to breath timing
control philosophies
1
Phi-
loso-
phy
Adult Pediatric Infant/
neona-
tal
(S)CMV P-CMV
APV
cmv
SIMV P-SIMV
APV
simv
(S)CMV
SIMV
P-CMV
P-SIMV
APV
cmv
APV
simv
P-CMV
P-SIMV
APV
cmv
APV
simv
I:E/
Pause
I:E,
Pause
I:E TI,
Pause
TI TI,
Pause
TI TI
Peak
Flow/
Tip
Peak
Flow,
Tip
%TI Peak
Flow,
Tip
%TI TI,
Pause
TI TI
%TI/
Pause
%TI,
Pause
%TI %TI,
Pause
%TI TI,
Pause
TI TI
TI/
Pause
TI,
Pause
TI TI,
Pause
TI TI,
Pause
TI TI
1. SPONT, ASV, DuoPAP, NIV, DuoPAP, and APRV modes are not shown, because breath control
timing philosophies do not apply to these modes.