9-8 624074/07
9 P/V tool maneuver option
Figure 9-3. P/V Tool window
1 Settings
1
. Shows the currently active P/V Tool control settings.
These settings are:
•
Pstart.
PEEP to be applied at start of maneuver.
•
Ptop.
Maximum pressure to be applied.
•
End PEEP.
PEEP to be applied at end of deflation limb.
If Pstart and End PEEP are different. A window for con-
firmation displays asking whether you want to change
PEEP/CPAP setting after maneuver.
•
Ramp speed.
Rate of change of pressure.
•
Tpause.
Length of pause between inflation and
deflation limbs.
2 •
Tmaneuver.
Total maneuver time, calculated.
•
Vpeep.
Released volume at the start of a ZEEP maneuver.