624074/07 9-9
3
Start/Stop.
Begins maneuver (or ends it at any time).
4 Lets you scroll through the previous stored PV Tool curves.
5 Plot number/Total number of stored plots
6
History.
Lets you scroll through the previous 5 stored
P/V Tool curves.
7
Cursor 1
and
Cursor 2.
Each cursor simultaneously selects
two points, one for the inflation and one for the deflation limb.
The points are vertically aligned. Using both cursors you can
determine the linear compliance by obtaining a “best visual fit.”
Data about these points is also displayed.
8 Cursor data. The
Cursor
column shows the values for the
points on the inflation limb (
Infl limb
) and deflation limb
(
Defl limb
) of the curve. The
Cursors
column shows the
calculated compliance in ml/cmH2O for a straight line that
connects the cursor points.
9 Curve, for this example, a pressure/volume curve that includes
both inflation and deflation limbs.
10 Time/date. Maneuver date and start time
1. See Table A-5 for a list of setting ranges, standard settings, and resolutions.