F-46 624074/07
F Pulse oximetry
Table F-12. Troubleshooting Masimo issues
Code or Issue Description Action
For details on Masimo information codes, see Section F. 7 .
No sensor informa-
tion in
System ->
Info 2
window
The ventilator does not recog-
nize the oximeter. This can
occur for several reasons:
• Masimo is not properly
enabled or configured.
•In
Configuration
,
ensure Masimo is
enabled. See Section
F. 8.1 .
• Ensure SpO
2
moni-
toring is enabled. See
Section F.8.1.3.
No sensor informa-
tion in
System ->
Info 2
window
• The SpO2 module is not
properly seated or
installed.
• Reinstall the module
per the installation
instructions.
• If the above solutions
do not address the
issue, contact a
HAMILTON MEDI-
CAL authorized ser-
vice technician.
Line 2:
MCU V1.1.0.6-
301-1-c-205f-0
where the last digit
is any number
from 1 to 10
Codes 1 to 10 indicate a
Masimo adapter board failure.
Contact a HAMILTON
MEDICAL authorized ser-
vice technician.
Line 2:
Sensor 3-3-6-581f-
00000000
where the last
code is any num-
ber other than
00000000.
Codes other than 00000000
indicate a Masimo sensor or
cable status or failure.
The possible issues are
described below.