624074/07 F-47
0000 0001 No cable connected. Connect cables.
The time allotted for the
cable is up. Replace the
cable. The information in
line 4 (Table F-8) always
shows the remaining
time or spot checks left.
0000 0002 Cable life expired. The time allotted for the
cable is up. Replace the
cable.
The information in line 4
(Table F-8) always shows
the remaining time or
spot checks left.
0000 0004 Incompatible cable. Replace with a compati-
ble Masimo cable.
0000 0008 Unrecognized cable. Reconnect the cable. If
this does not resolve the
issue, replace the cable.
0000 0010 Defective cable. Replace the cable.
0000 0080 No sensor connected. Connect a Masimo sen-
sor.
0000 0100 Sensor life expired. The sensor has reached
the end of its allotted
hours or spot checks.
Replace the sensor.
The information in line 4
(Table F-8) always shows
the remaining time or
spot checks left.
0000 0200 Incompatible sensor. Connect a compatible
Masimo sensor.
Table F-12. Troubleshooting Masimo issues
Code or Issue Description Action
For details on Masimo information codes, see Section F. 7.