624074/07 A-7
IntelliCuff
®
Minimum pres-
sure 0-50 cm
H2O
--
Manual mode 5-50 cm H2O 20 cmH2O --
Automatic mode
Rel. pressure -15 to 5 cmH2O 0 cmH2O --
Min. pressure -15 to
50 cmH2O
20 cmH2O --
Max. pressure 5 to 50 cmH2O 30 cmH2O --
Flowtrigger
Adult 0.5 to 15 l/min 5 l/min 0.5 l/min
Pediatric 0.5 to 15 l/min 3 l/min 0.5 l/min
Infant/neonatal 0.1 to 5 l/min 1.5 l/min 0.1 l/min for
< 2.0 l/min
0.5 l/min for
≥
2.0 l/min
Gender Male, female Male --
I:E
1
Adult 1:9 to 4:1 1:2 1 for 1:9 to 1:4
0.1 for 1:4 to 4:1
Pediatric/infant/
neonatal
-- -- --
Loudness (alarm) 1 to 10 5 1
%MinVol (% minute volume)
Adult/pediatric 25 to 350% 100% 5%
Infant/neonatal -- -- --
Table A-5. Control setting ranges and resolutions
Setting Range Standard
setting
1
Resolution