624074/07 A-27
ExpMinVol
low
l/min Off
1
0.1-49
4- ASV:
(IBW*0.1-
40%)*%Min-
Vol, if >100%
- NIV, NIV-ST,
nCPAP-PS:
ExpMinVol -
70%
- All other
modes:
ExpMinVol -
40%
1
Pulse rate
high
bpm 5 - 235 in incre-
ments of 5
140 Same as stan-
dard settings
1
Pulse rate
low
bpm 30 - 230 in
increments of 5
50 Same as stan-
dard settings
1
Pressure
high
cmH2O 10-120 40 - Volume
modes/ASV:
Ppeak+
10 cmH2O or
minimum of
40 cmH2O
- APV: Ppeak+
10 cmH2O
- Pressure
modes:
Pcon-
trol+PEEP+
10 mbar
- SPONT:
Psupp+
PEEP+10 mbar
- NIV, NIV-ST,
nCPAP-PS:
2*Psupp+PEEP
+10 mbar
1
Pressure
low
cmH2O 2-119 5 PEEP+5 mbar 1
Table A-10. Adjustable alarms with automatic and standard settings
Parameter Unit Range Standard
setting
Automatic
setting
Resolu-
tion