A-28 624074/07
A Specifications
Vt high ml Off,
100-3000
750 Vt+50% <1000:
10
>1000:
50
Vt low ml Off,
50-2950
250 Vt-50% <1000:
10
>1000:
50
Rate high b/min 2-130 23 - Volume
modes/ASV:
Rate/fCon-
trol+40%
- All other
modes:
fTotal+40%
Rate low b/min 0-128 8 - Volume
modes/ASV:
Rate/fControl-
40%
- All other
modes:
fTotal-40%
Apnea time s 15-60 20 Same as stan-
dard settings
5
VLeak high % Off,
5-80%
Off -- 5%
EtCO
2
high mmHg
Torr
Off,
1-100
60 EtCO
2
+25% 1
EtCO
2
high kPa Off,
0-13.3
8EtCO
2
+25% 0.1
EtCO
2
low mmHg
Torr
Off,
1-99
30 EtCO
2
-25% 1
EtCO
2
low kPa Off,
0-13.2
4EtCO
2
-25% 0.1
Table A-10. Adjustable alarms with automatic and standard settings
Parameter Unit Range Standard
setting
Automatic
setting
Resolu-
tion