624074/07 A-29
Oxygen
high
% Oxygen +5%
(permanently,
not with
“Auto” Button
Oxygen low % Oxygen -5%
(permanently,
not with
“Auto” Button
PI high % 0.04 - 0.10, in
increments of
0.01
0.10 - 1.0, in
increments of
0.10
1.0 - 19, in
increments of 1
off Same as stan-
dard settings
0.01
0.1
1
PI low % 0.03 - 0.10, in
increments of
0.01
0.10 - 1.0, in
increments of
0.10
1.0 - 18, in
increments of 1
off Same as stan-
dard settings
0.01
0.1
1
SpCO high % 2 - 98, in incre-
ments of 1
10 Same as stan-
dard settings
1
SpCO low % 1 - 97, in incre-
ments of 1
off Same as stan-
dard settings
1
SpO
2
high % Off
71-100 (Nihon
Koden)
2-99 (Masimo)
99 Same as stan-
dard settings
1
SpO
2
low % 85-99 (Nihon-
Kohden)
90 Same as stan-
dard settings
1
Table A-10. Adjustable alarms with automatic and standard settings
Parameter Unit Range Standard
setting
Automatic
setting
Resolu-
tion