624074/07 3-21
1. Connect the nebulizer unit to the T-piece by pushing the
nebulizer unit firmly into the T-piece.
Figure 3-12. Connecting nebulizer to T-adapter
2. Insert the nebulizer and the T-piece into the breathing cir-
cuit with the arrow on the T-piece pointing in the direction
of the air-flow within the circuit.
3. Connect the Aeroneb nebulizer to the Aerogen module
using the nebulizer cable.
Figure 3-13. Connecting HAMILTON-G5 Aeroneb module