624074/07 5-29
• You can selectively enable or disable some adjustable
alarms during configuration (see Appendix J, Config-
uration). At the minimum, the high pressure, low
ExpMinVol, and apnea time alarms are always active.
• The number to the left of the line through each
alarm column is the current measured value. The
circled numbers at the top and bottom of the alarm
columns represent the alarm settings.
• After a power interruption of up to 30 seconds,
the device restores the last settings, including alarm
settings.
You can access the
Alarms
window and change alarm settings
at any time, without affecting ventilation.
The device offers two alarm-setting options:
• Manually set individual alarm limits.
•Use the
Auto
alarm function. Choosing
Auto
sets all alarm
limits. Table 5-3 lists the Auto alarm settings.
To review and adjust alarms:
1. Open the
Alarms
window (Figure 5-10). The
Limits 1
window is displayed.
Figure 5-10. Alarm Limits 1 window
If the Pulse Rate alarm and/or one or more of the Masimo
parameters is enabled in configuration, the
Alarms ->
Limits 2
window is active. Select the
Limits 2
tab to
display the window.