EasyManua.ls Logo

Hamilton G5 - Page 203

Hamilton G5
594 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
624074/07 5-33
PetCO
2
(low
and high)
Low and high monitored PetCO
2
.
If either limit is reached, a medium-
priority alarm is annunciated.
Low: Off, 0 to
99 mmHg/0 to
13.2 kPa
High: Off, 1 to
100 mmHg/1
to 13.3 kPa
Pressure (low
and high)
Low and high monitored pressures at
the patient airway (Ppeak). If either
limit is reached, a high-priority alarm is
annunciated.
In addition, if the high Pressure limit is
reached, the HAMILTON-G5 immedi-
ately stops gas flow to the patient and
relieves pressure until pressure falls to
the PEEP/CPAP level. If the airway pres-
sure continues to rise above
120 cmH
2
O, the mechanical overpres-
sure relief valve opens and the ventila-
tor enters the ambient state.
Low: 2 to
119 cmH
2
O
High: 10 to
120 cmH
2
O
The HAMILTON-G5 ventilator uses the
high Pressure limit as a safety boundary
for its inspiratory pressure adjustment.
ASV
®
, INTELLiVENT
®
-ASV, APV-cmv
and APV-simv do not apply inspiratory
pressures higher than 10 cmH
2
O
below the high Pressure limit. An
exception is sigh breaths, when ASV
®
,
INTELLiVENT
®
-ASV, APV-cmv and APV-
simv can apply inspiratory pressure
3cmH
2
O below the high Pressure
limit.
In the ASV
®
and INTELLiVENT
®
-ASV
controls window the high Pressure
alarm can also be changed by adjust-
ing the Pasv limit (maximally applicable
pressure target 10 cmH
2
O below high
Pressure alarm limit).
Table 5-3. Alarm limit settings and ranges
Parameter Definition Range

Table of Contents

Related product manuals