624074/07 7-23
MV Spont
NIV
(ml)
Monitoring 1 Spontaneous expiratory minute volume. The
moving average of the monitored expiratory
volume per minute for spontaneous breaths,
over the last 8 mandatory and spontaneous
breaths.
Used only in Noninvasive modes (NIV, NIV-ST,
nCPAP-PS), this parameter replaces MV Spont,
as it is leak compensated.
Oxygen Monitoring 1 Oxygen concentration of the delivered gas. It is
measured by the oxygen cell in the inspiratory
pneumatics.
This parameter does not display when: a) the
oxygen supply is not connected, b) the oxygen
cell is not installed or is defective, or c) if oxy-
gen measurement is disabled.
Table 7-1. Monitored parameters