Monitoring 1 Inspiratory pressure time product. The mea-
sured pressure drop required to trigger the
breath multiplied by the time interval until the
PEEP/CPAP level is reached at the beginning of
inspiration.
Both PTP and WOBimp (see below) indicate
work by the patient to trigger the breath. The
work depends on
the intensity of the patient’s effort,
the trigger type and sensitivity, and
the volume and resistance of the breathing
circuit.
PTP and WOBimp are valid for patient-initiated
breaths only.
Neither PTP nor WOBimp indicates total patient
work. But both indicates of how well the venti-
lator is adapted to the patient.
If PTP and WOBimp values increase:
Check and remove water in tubes.
Switch to flow trigger.
Increase trigger sensitivity.
Pulse
(1/min)
Monitoring 2 Pulse rate of the patient measured by the SpO
2
sensor.
Pulse inc
(%)
Monitoring 2 Quick Wean parameter. Percentage increase in
heart rate. This is calculated during a spontane-
ous breathing trial (SBT).
For details, see the Quick Wean chapter in the
INTELLiVENT
®
-ASV manual.
QI-HLI Monitoring 2 Quality Index of the HLI reflects the reliability of
the HLI value expressed in percentage. The
greater the QI value the greater reliability of the
HLI value.