624074/07 10-11
ASV: Can-
not meet
target
Low priority.
The operator-set %MinVol can-
not be delivered, possibly
because of setting conflicts.
Observe the patient.
Verify the control settings.
Consider decreasing the
%MinVol setting or increas-
ing the high Pressure alarm
limit to an appropriate
level.
Consider suctioning or
other therapy.
NOTE:
Display the ASV
target graphics screen
to help troubleshoot
this alarm.
ASV: Check
high pres-
sure limit
Low priority.
The operator-set high pressure
alarm limit is too low, and the
ventilator cannot deliver the cal-
culated target tidal volume.
Observe the patient. Con-
sider suctioning or other
therapy.
Verify the control settings.
Consider increasing the
high Pressure limit to an
appropriate level.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions