624074/07 10-31
SBT suc-
cessfully
fulfilled
Medium priority. The SBT was
ended because Duration max
was reached.
None.
Touch the
I-icon
or the
alarm message to dismiss
the alarm.
Screen lock
active!
Operator message.
Screen lock key is pressed.
Unlock screen if user entry
is desired.
Serial num-
ber doesn’t
match!
Operator message.
Entered option key is for a device
with a different serial number.
Contact HAMILTON MEDI-
CAL representative:
SpCO high Low priority.
SpCO value measured by the
sensor exceeds the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SpCO low Low priority.
SpCO value measured by the
sensor is below the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SpHb high Low priority.
SpHb value measured by the sen-
sor exceeds the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SbHb low Low priority.
SpHb value measured by the sen-
sor is below the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SpMet high Low priority.
SpMet value measured by the
sensor exceeds the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SpMet low Low priority.
SpMet value measured by the
sensor is below the set limit.
Observe the patient. Verify
ventilator settings, includ-
ing alarm settings.
SpO
2
: sen-
sor error
(left slot)
Medium priority.
Hardware problem with
sensor adapter
Replace left sensor adapter.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions