624074/07 10-33
SpO
2
left
sensor light
interfer-
ence
Operator Message. Light interfer-
ence in left sensor probe. Second
SpO
2
sensor still working nor-
mally.
Cover left probe with blan-
ket or change attachment
site on patient. Replace left
probe.
SpO
2
: poor
signal (left
slot)
Operator Message. Pulse from
left SpO
2
sensor not found. Sec-
ond SpO
2
sensor still working
normally.
Verify left probe attach-
ment, check if left probe
site is properly perfused.
SpO
2
: sen-
sor error
(right slot)
Medium priority.
Hardware problem with
sensor adapter.
Replace right sensor
adapter.
SpO
2
:
patient dis-
connected
(right slot)
Medium priority.
Probe malfunction. Probe is
detached from patient or not
properly attached to patient.
Verify if right probe is
attached properly to
patient.
Replace right probe.
SpO
2
: no
sensor
(right slot)
Medium priority.
Probe is disconnected from
adapter. Adapter is disconnected
from ventilator. Cable defective.
Connect right probe to
adapter.
Connect right sensor
adapter to ventilator.
Replace right adapter and/
or probe.
SpO
2
: light
interfer-
ence (right
slot)
Medium priority.
Light interference in sensor probe
Cover right probe with
blanket or change attach-
ment site on patient.
Replace right probe.
SpO
2
: poor
signal (right
slot)
Medium priority.
Pulse from right SpO
2
sensor not
found.
Verify right probe attach-
ment, check if right probe
site is properly perfused
SpO
2
right
sensor
adapter
faulty
Operator Message. Hardware
problem with right sensor
adapter. Second SpO
2
sensor still
working normally.
Replace right sensor
adapter.
Table 10-2. Alarms and other messages
Alarm Definition Complete these actions