624074/07 11-5
Mask (silicone rubber) Steam autoclave,
chemically disinfect,
or ETO sterilize
Avoid exposing silicone
rubber masks to grease,
oil, silicone-based lubri-
cants, organic solvents
(benzene, ether, ketone,
and chlorinated hydro-
carbons), acids, concen-
trated alkaline cleaning
products, and phenols
and derivatives.
Deflate air cushion before
steam autoclaving to
prevent possibility of
explosion.
Flow sensor, single-use Chemically disinfect (only
if this single-use Sensor
requires decontamina-
tion before use)
The flow sensor is
designed for single-
patient use. It is delivered
clean and ready for
patient use. If the sensor
must be decontaminated,
DO NOT use hard
brushes, pointed instru-
ments, or rough mate-
rials. These can damage
the flow sensor’s mem-
brane.
Flow sensor, reusable Chemically disinfect or
ETO sterilize
DO NOT use hard
brushes, pointed instru-
ments, or rough mate-
rials. These can damage
the flow sensor’s mem-
brane.
Inspiratory filter,
reusable autoclavable
Steam autoclave Always inspect the filter
media for cracks or for-
eign matter; replace if
necessary. Replace after
20 autoclave cycles.
NEVER chemically disin-
fect or expose to ETO gas.
Table 11-1. Decontamination methods for HAMILTON-G5 parts
Part (material) How to decontaminate Remarks