624074/07 11-7
CO
2
sensor Clean and disinfect the
outside by wiping with
a cloth dampened with
70% isopropyl alcohol,
a 10% aqueous solution
of sodium hypochlorite
(bleach), disinfectant
spray cleaner such as
Steris Coverage® Spray
HB, ammonia, or mild
soap.
Wipe down with a clean
water-dampened cloth to
rinse, and dry before use.
Make sure the sensor
windows are clean and
dry before reuse.
Always disconnect the
CO
2
sensor before clean-
ing. DO NOT immerse or
attempt to sterilize the
sensor.
Before reusing the sensor,
ensure the windows are
dry and residue-free, and
that the sensor has not
been damaged during
handling or by the
cleaning process.
Replace if damaged or if
excessive secretions are
observed.
WARNING
Reusing, disassembling, cleaning, disin-
fecting, or sterilizing the single-use CO
2
airway adapter can compromise its func-
tionality and system performance, leading
to a possible user or patient hazard. Per-
formance is not guaranteed if an item
labeled as single-use is reused.
Table 11-1. Decontamination methods for HAMILTON-G5 parts
Part (material) How to decontaminate Remarks