624074/07 11-9
SpO
2
Adapter Periodically clean the
SpO
2
adapter. Wipe with
a soft cloth moistened
with ethanol (15°C
(59°F), 76.9 to 81.4%
by vol). Dry the SpO
2
adapter completely after
cleaning.
After cleaning before use,
wipe liquid off with a dry
cloth and thoroughly dry
the SpO
2
adapter.
DO NOT immerse the
SpO
2
adapter in chemical
solution or water.
If the SpO
2
adapter is
immersed, wipe off liquid
with a dry cloth and
thoroughly dry the SpO
2
adapter before use.
DO NOT disinfect and
sterilize the SpO
2
adapter.
It will damage the SpO
2
adapter.
SpO
2
probe Refer to the operator’s
manual of the probe you
use.
Aeroneb adapter Autoclave wrapped parts
using steam sterilization
pre-vacuum cycle, a mini-
mum of 134°C (270°F –
275°F) for 20 minutes
with drying cycle (some-
times referred to as
a “Prion cycle”).
DO NOT reassemble parts
prior to autoclaving.
Aeroneb control mod-
ule, control module
cable and AC/DC
adapter
Wipe clean with a damp
cloth. DO NOT autoclave.
Check for exposed wir-
ing, damaged connec-
tors, or other defects and
replace if any are visible.
Aeroneb mounting
brackets
Wipe clean with a damp
cloth and mild liquid
detergent. DO NOT use
abrasive or sharp tools.
Table 11-1. Decontamination methods for HAMILTON-G5 parts
Part (material) How to decontaminate Remarks