624074/07 A-31
PEDIATRIC (only where differing from ADULT values)
ExpMinVol
high
l/min Off,
0.3-10
3.5 - ASV:
(IBW*0.2+
40%)*%Min-
Vol, if >100%
- All other
modes:
ExpMinVol +
40%
1
ExpMinVol
low
l/min Off
2
0.1-9.8
1.5 - ASV:
(IBW*0.2-
40%)*%Min-
Vol, if >100%
- NIV, NIV-ST,
nCPAP-PS:
ExpMinVol -
60%
- All other
modes:
ExpMinVol -
40%
1
Vt high ml Off,
10-500
150 Vt+50% <
100:10
>
100:10
Vt low ml Off,
0-300
50 Vt-50% <
100:10
>
100:10
Rate high b/min 2-130 38 - Volume
modes/ASV:
Rate/fCon-
trol+40%
- All other
modes:
fTotal+40%
Table A-10. Adjustable alarms with automatic and standard settings