A-44 624074/07
A Specifications
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
4-8
3 A/m 30 A/m The power frequency mag-
netic field should be at levels
consistent of a typical location
in a typical commercial or
hospital environment.
Portable and mobile RF com-
munications equipment should
be used no closer to any part
of the HAMILTON-G5 ventila-
tor, including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
Conducted
RF
IEC 61 000-
4-6
3 Vrms
150 kHz to
80 MHz
outside ISM
bands
2
10 Vrms
150 kHz to
80 MHz in
ISM bands4
10 Vrms
10 Vrms
d = 0.35 √P
d = 1.2 √ =P
Table A-15. Guidance and manufacturer's declaration – electro-
magnetic immunity
1
Immunity
test
IEC 60601
test level
Compli-
ance level
Electromagnetic environ-
ment – guidance