1-18 624074/07
1 General information
3 Press-and-turn (P&T) knob. Selects and adjusts ventilator
settings and selects monitored data.
4 Print screen key. Saves a JPG file of the current ventilator
screen to a CompactFlash or USB storage device.
5 Standby key. Activates the standby (waiting) mode. When the
mode is activated, the standby activated screen is displayed. For
details, see Section 12.1.
6 Nebulizer on/off key. Activates nebulization during the
breath phases and for the duration selected during configu-
ration. You can switch nebulization off earlier by pressing the
key again.
7 Screen lock/unlock key. Prevents inadvertent touch screen
entries. By touching the locked screen an acoustic BEEP sounds
and a notice “screenlock active” appears.
8 Alarm silence key. Silences the ventilator audible alarm for
2 min. Pushing a second time cancels the alarm silence.
9 O2 enrichment key/suctioning tool.
Adults and Pediatric: Delivers 100% oxygen for 2 min. The
backlighting changes color to green and the actually applied
oxygen concentration is displayed on the oxygen control
(green). Pushing the key a second time or manually changing
the oxygen concentration ends the 100% oxygen enrichment
period.
Neonatal option:
1
Delivers 125% of the last oxygen setting for
2 min. The backlit color changes to green and the currently
applied oxygen concentration is displayed on the oxygen knob
(green). Pushing the key a second time or manually changing
the oxygen concentration ends the oxygen enrichment period.
For further information on the suctioning tool, see
Section 12.3.
10 Manual breath key. Triggers a mandatory breath when
pressed and released during exhalation. The mandatory breath
is delivered using the currently active settings.
1. In Japan, 100% oxygen is also applied for the neonatal option