624074/07 F-51
Environmental
Operating tempera-
ture
0 to 50°C (32 to 122°F)
Storage temperature -40 to 70°C (-40 to 158°F)
Relative storage
humidity
10 to 95% noncondensing
Operating altitude pressure: 500 to 1060 mbar
Altitude: -304.5 to 486 m (-1000 to 18,000 ft)
Mode & sensitivity
SpO
2
averaging
mode
2, 4, 6, 8, 10, 12, and 16 seconds; FastSat
SpO
2
sensitivity APOD, Normal, Maximum
Alarms
Volume level adjust-
ment: pulse/tone
OFF; 25 to 100% in four increments
Alarm silence 120 second delay; All mute: continuous silence
Out of Limit alarms:
SpO
2
, Pulse rate,
SpCO, SpMet, SpHb,
PI, PVI
High/low alarms
Sensor condition
alarm
No Sensor; Sensor Off; Sensor Defect
Battery alarm Low battery
Table F-13. Masimo pulse oximeter specifications
Feature Specifications