624074/07 J-19
3. Define your defaults for the selected patient group.
A.In the
Modes
window, select the mode you want to set as
default for the active patient group.
B.Adjust all controls in the
Controls
window according to
your institution’s standards as default controls.
C.In the
Additions
window, set
TRC
and
Sigh
according
to your institution’s standards.
D.Use the graphics layout in the
Graphics
window to con-
figure the graphics layout. Activate the secondary moni-
toring parameters (
SMP
) according to your institution’s
standard for waveforms, loops, trends and the Intelli-
gent Panels for the default graphics layout.
4. After setting mode, controls, addition and graphics return
to the
Standby
window and activate the
Configuration
> Defaults
window.
For the last active patient group (where you set your
defaults) the
Set default
button is available.
5. Save your default configuration for the active patient
group. After pressing the
Set default
button a confirma-
tion window opens.
6. Repeat these steps for the remaining patient groups.
The default patient group can also be selected in the
Defaults
window. This patient group will be the default
for
New Patient
.
7. Select the current patient group. A drop-down list opens.
8. Touch
Close
.
9. If you are done configuring the ventilator, touch
Close/
Save
to save your selections.
Note that selecting
Set factory defaults
deletes all user-
configured defaults (mode, controls and graphics).